1. Press and hold down the operator button for four (4) seconds
2. The screen will display CAL and the O2 icon will flash in the lower left hand
side
3. After a successful calibration, the personal monitor will emit one (1) beep,
vibration and LED flash
4. After an unsuccessful calibration, the personal monitor will beep 5 times,
flash, and continue to display calibration. If after a few failed calibrations
please contact Crowcon customer support or your local service centre.
6. Hibernation Instructions – Crowcon Clip+ only
Crowcon Clip+ Hibernation via the IR Link
1. Check that you have installed the IR Link Software and the IR Link USB connections are
plugged in
2. Click Read Device on the IR Link Software
3. Note; when the monitor is hibernated the event log will be cleared. It is highly recommended
to save the event log by pressing the save event log before hibernating.
4. Click on the Hibernate Button, acknowledge the event log message
5. Keepthe Clip+ in front of the IR Link until the "Hibernate OK" message is displayed at the
bottom of the IR Link Software
6. Confirm the monitor screen is blank
7. If you encounter any problems please contact Crowcon or your local service centre.
Crowcon Clip+ Hibernation via C-Test
1. Check that the C-Test is turned on and the USB memory stick is present
2. The Clip C-Test is capable of hibernating the Crowcon Clip+, place the monitor in the C-
Test
3. Press and hold the bump test and calibration buttons down simultaneously for approx 2
seconds
4. A successful hibernation will result in a GREEN light on for the C-Test
5. Note the event log will be automatically stored on the C-Test USB memory
6. Confirm the monitor screen is blank
7. If you encounter any problems please contact Crowcon or your local service centre
Issue 4 – December 2016
Page 9