Argos Technologies EVac Manual do operador - Página 16

Procurar online ou descarregar pdf Manual do operador para Equipamento de laboratório Argos Technologies EVac. Argos Technologies EVac 20 páginas.

Argos Technologies EVac™ — Operator's Manual

Troubleshooting (FAQ)

Problem
Equipment not
running; no
vacuum
Suction power
inadequate
or pump runs
constantly
Alarm level
detection
Possible Cause
No power line voltage
No DC supply
Pump switched off
Suction power selected
too low
Leak in vacuum system
Bottle full; liquid
sucked into filter
Tubes Blocked
Filter blocked
Sensor cable
incorrectly connected
Medium or foam in
bottle has reached
maximum level
Level detection not
calibrated
Incorrect calibration
(Four beeps will sound)
The lid interior between
sensor pins is soiled
15
Remedy
Insert mains adapter plug into
suitable power outlet.
Connect mains adapter to
instrument.
Turn power switch to "ON"
(l). Ensure that LED indicator
under bottle is lit.
Turn control knob clockwise.
Check tubes and filters.
Check lid and tighten if
necessary.
Check if sealing plug/venting
screw on bottle lid is leaking.
Ensure that filter is fully
tightened onto unit.
Ensure that leak flow of hand
tool is not excessively high.
Remove wet filter and replace
with new, dry filter.
Empty excess liquid from
existing bottle or replace with
new, empty bottle.
Clean or replace tubes.
Replace filter.
Check that cable is correctly
connected to device and lid.
Empty excess liquid from
bottle. Use an anti-foam agent.
Calibrate level detection
(see page 13).
Check sensor pins, sensor
cable and connector. Retry
calibration (see page 13).
Clean and dry interior of lid
between sensor pins.