MSI MSI-8000HD Manual do operador - Página 25

Procurar online ou descarregar pdf Manual do operador para Monitor MSI MSI-8000HD. MSI MSI-8000HD 48 páginas. Rf remote display

MSI MSI-8000HD Manual do operador

5.1 RF Setup

RF Setup Procedure
1) With the power on, open the
Comm Setup Menu by
pressing the
F1
simultaneously.
The COMM Setup Menu
appears. The first menu
choice is "Print".
2) Press
F2
to scroll to the RF
Setup menus.
3) Press
F1
to enter the RF
setup menu.
4) The display reads "OnOff"
Confirm the 8000 RF is on
by pressing F1. If it is off,
use the F2 key to change it
to on. Then push F1.
Off is only used when the 8000 is
hardwired to a Dyna-Link.
5) Press F1. Next the "Scid"
(ScaleCore ID) screen is
shown
6) Press F1. The current Scid
number is shown.
If the offered Scid value is correct,
push
F1
and jump to step 10.
In this example we'll enter 28 as an
SCID number. Any value from 1-254
is acceptable. However, MSI
recommends a value from 20 – 30.
7) To input the SCID value,
press
F2
to start the number
entry process.
8) Use
F2
to change the
number, and
the number. Add the next
digit by pushing
scrolling as required.
Repeat this sequence
until the entire SCID
number is entered.
9) Once the entire SCID
value is entered, press
to finalize the number.
10) The next RF Setup
parameter is "ChnL"
(Channel). Press
the number entry process.
Press Simultaneously
and
F3
keys
SCROLL
ENTER/SELECT
ENTER/SELECT
ENTER/SELECT
ENTER/SELECT
SCROLL
ENTER/SELECT
F1
to enter
F2
and
Error Correction: If you input a wrong value, press
back one digit and reenter.
F1
fixed
F1
to start
Figure 5-2. RF Set Up
example ScID Value
blinking
fixed
SCROLL
F2
8 times
ENTER/SELECT
ENTER/SELECT
Exit and Cancel Changes
Exit and Save/Back 1 Level
0
F1
SCROLL
SCROLL
ENTER/SELECT
blinking
ZERO
COMM Set Up
Enter / Select
Scroll
F2
blinking
blinking
to step
21