Choosing The Working
Space For The Unit
Care should be taken when choosing the working space for this unit.
The unit should be used indoors only, in a cool
dry room with stable temperature and humidi-
ty, kept out of direct sunlight, away from any
heat sources, excess moisture or naked
flames.
This unit should only be operated in a
temperature range of 10° to 40° C (50° to 103°
F). It should not be placed in the direct path of
a room air conditioning outlet.
The full unit weighs 300 kg (660lbs); it should
be placed on a level stable floor.
It should never be placed on a soft or uneven
surface like bedding materials or shaggy
carpets.
The unit should be placed in a well-ventilated
position, and all inlets and outlets should be
unobstructed, there should be a 15cm or
6-inch clearance on all sides.
≥15cm
04
Parameters
Professional Veterinary
Product Name
Intensive Care Unit
Model Number
UC-1901M/L/BK
1509mmW×884mmD×1899mmH
Dimension
(59.5"W×35"D ×75"H)
Approximate 300kg (660lbs)
Weight
AC100~120V /220~240V
Power Supply
50/60Hz (Grounded)
4 Compartments Controlled
Separately, Cooling and Heating.
Two Medium Compartments on
Environment
Top Tier, One Large Compartment
Control
on Bottom Tier which could be
divided into two medium com-
partments when needed.
Ultra Violet Sterilization, High
Air Purification
Concentration Negative Ion Air
Purification.
Working
-10~40°C (Indoor)
Condition
Temperature
15~38°C, Tolerance at 0.5°C
Setting Range
Oxygen Concen-
21~65%, Tolerance at ±1%.
tration Setting
Carbon Dioxide
200 -5000PPM,
Monitoring and
Tolerance at ±10PPM,
Removal
Built-in CO2 Sensor and Remover
O2 Concentration Alarm, Tempe-
rature Alarm, Sensor Alarm, Door
Opening Alarm, CO
2
Alarms
Concentration.
Alarm, Emergency Ventilation Port
Malfunction Alarms.
Automatic Dehumidification,
High/Middle/Low Three Stages,
Dehumidification
High: 45%~50%,
Middle: 40%~44%,
Low: 30%~39%.
Control Panel
8.4", 171mmW×128mmH
Screen Type
Resistive
Max. Power
1.3KW
Consumption
Average Power
0.5KW
Consumption
Power Supply
Emergency Ventilation Port
Failure Protection
Negative Ion
7.2×106pcs/cm3
Generator
Negative Ion Generator
Soda Lime CO2 Absorption
Accessories
Powder, Equipment Mount Grid,
Door Shade Curtain.
Built-in Printer
Smart Control Needle Type Printer
Ultra-violet
253nm UV Sterilization
Sterilization
Blue Light
415nm Blue Therapeutic Light
Therapy
Treatment
Medical
≥0.2ml/min, particle size:
Nebulizer
0.5 -2μm, noise≤ 40dB(A)
Pure White Cold Light for
Examination, 10 Stage Adjustable
Yellow .
LED Illumination
Comfort Light for Illumination.
Low/Middle/High, Three-Stage Speed
Control.
Defaults:
Oxygen Input On: Low Speed.
Ventilation Fan
Temperature Control: High Speed
Control
Emergency Port Open: High Speed
Ultra-violet Sterilization On: Middle
CO2 Concentration Level Alarm On:
High
05