In this transmission mode the instrument transmits continuously the following data format depending on the functioning
mode:
Totalisation Mode
ST,GS,XXX.XXXXkg<CR><LF>
<CR><LF>
Counting Mode
PCS:XXXXXXXXpcs<CR><LF>
<CR><LF>
Percentage Mode
PERC:XXXXX.XX%<CR><LF>
<CR><LF>

9. ERROR MESSAGES

ERROR MESSAGES
- - - - - -
Err 4
Err 5
Err 6

RECYCLING INSTRUCTION

The crossed-out wheeled bin on the product means that at the product end of life, it must be taken to
separate collection or to the reseller when a new equivalent type of equipment is purchased. The
adequate differentiated refuse collection in having the product recycled, helps to avoid possible
negative effects on the environment and health and supports the recycling of the materials of which
the equipment is made. The unlawful disposal of the product by the user will entail fines foreseen by
the current regulations.
GS for gross weight, NT for net weight, XXX.XXXX is the weight
XXX.XXXX is the Number of pieces
XXXXX.XX is the Percentage of weight
DESCRIPTION
RESOLUTION
Over range
Remove weight from the scale.
Zero Setting Error
Upon start-up or when the ZERO key is pressed, the weight on the
scale exceeds the percentage programmed on the max. capacity.
Remove the excess weight and retry.
Keyboard Error.
The keyboard could be damaged.
Input signal greater
Indicates that the A/D converter value is greater than the maximum
than 3mV/V
value:
- Remove the weight from the scale if it's exceeding.
- The load cell or the electronics could be damaged.