Dini Argeo DGT4XDEVNET Manual de início rápido - Página 9
Procurar online ou descarregar pdf Manual de início rápido para Transmissor Dini Argeo DGT4XDEVNET. Dini Argeo DGT4XDEVNET 17 páginas. Digital weight transmitter with 4 channels

13. RS485 port
888888
div.deC
...
15
485
i
The RS485 port is configured by default to communicate in Modbus RTU (ch. 18).
14. Broken laod cell exclusion (for dependent channels systems)
If a load cell is broken, it's possible to temporarely exclude the channel where it is connected and continue to weigh, pending
replacement.
WARNING: this operation reduces the accuracy of the weighing system. We recommend use for liquid weighing or in applications
where the load is evenly distributed.
888888
div.deC
...
16
exCl.Ch
Visible only in DEP.Ch mode.
Complete menu
at page 4
1
485 address (01 ÷ 98).
se.add
div.deC
2
Baud rate (1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200).
bad
1
No channel excluded.
none
div.deC
2
Ch 1
Channel 1 excluded.
3
Channel 2 excluded.
Ch 2
4
Channel 3 excluded.
Ch 3
5
Ch 4
Channel 4 excluded.
QSG_ENG_DGT4X
9