INTRODUCTION

The program is designed to perform the calibration of material testing equipment, machines
force
samples,
test
benches
or
other
equipment
using
standard
devices.
Evaluation of the uncertainty of calibration is performed according to the requirements of the
UNI CEI ENV 13005 and the information given in Appendix D of ISO / DIS 7500-1.
To perform the calibration is necessary to generate a significant number of measures (5, 8 or
10) of increasing strength in two or three series to check the repeatability of the machine
calibration.
After each cycle, increasingly, the machine must be reset to zero before performing the next
measurement cycle.
After the test, the program calculates:
a) Average readings
b) Error reading%
c)% expanded uncertainty associated with each point of measurement.
All certificates are then printed and stored in a database that keeps the historian of
calibrations performed.
The program maintains a database of standard AEP instruments type TA2USB , TAUSB,
MP10, DFI MP2E, MP2000A STAR, MP10Plus and MP6A connected to a PC via RS232
serial communication or USB to ensure the acquisition of instantaneous force measurement.
For each standard instrument is necessary to introduce all the identification data, SIT
certificates or equivalent
reference and for the different calibration points the computed
uncertainties.
The program automatically advices the operator if the used reference instrument has a
certificate date expired.