 mV / V: accessible only with the instrument TAUSB. Setting mandatory to calibrate the
sensitivity of the instrument with the sensitivity of output of the load cell connected.
Setting range: from 5000 (0.5mV) to 25000 (2.5mV), default value 2 (2mV). For
instruments other than TAUSB this field will be empty.
 Channel : accessible only with the instrument TA2USB. It allows to set the channel of
the TA2USB used during calibration
 Uncertainty Compression: In this table, the operator must introduce in the first
column the calibration points on the certificate SIT in load unit, and in the column, the
uncertainty associated with each point of strength in compression.
The table can accept up to 10 different strengths, but if necessary the operator can set
5, 7 or more depending on the certificate in his possession.
 Tensile Uncertainty : In this table, the operator must introduce in the first column the
calibration points on the certificate SIT in load unit, and in the column, the uncertainty
associated with each point of strength in tension. The calibration points should be
placed in tensile with a minus sign (-) (eg "-2.5").
The table can accept up to 10 different strengths, but if necessary the operator can set
5, 7 or more depending on the certificate in his possession.
WARNING:
The introduction of incorrect data affect the calculation of the uncertainty on the machines in
calibration.