Troubleshooting
Fault
No gas flow
Audible gas leak
Constant gas flow
Probes (quick connectors) for use with medical gas pipeline systems
Medical gas systems. Low pressure flexible hose assemblies
Possible Cause
Flowmeter is not connected
properly.
Gas cylinder empty.
Medical gas terminal unit on a
pipeline system is isolated.
Inlet filter blocked.
Flowsafe™ Valve has activated.
Firesafe™ Nozzle has activated
due to fire or mechanical damage.
Seal is worn or damaged.
Flow tube cracked or damaged.
Probe nose worn or damaged.
Medical gas hose damaged.
Valve or valve seat damaged.
Check gas supply. Check that the gas specific probe
is correctly connected.
Replace gas cylinder.
Seek advice from someone authorised to operate the
medical gas pipeline system isolation valves.
Repair required.
Flowsafe™ Valve is designed to activate when an
exceptionally high flow of gas passes through the
device. Disconnect from the gas supply and check the
flow meter for damage or rupture.
Replace Firesafe™ Nozzle.
Repair required.
Repair required.
Repair required.
Replacement required.
Repair required.
9
Solution