Calea Ambix Activ Manual do utilizador - Página 12

Procurar online ou descarregar pdf Manual do utilizador para Bomba de água Calea Ambix Activ. Calea Ambix Activ 16 páginas. Ambulatory pump & accessories
Também para Calea Ambix Activ: Manual de referência rápida (2 páginas)

Ambix Act iv Ambula t ory Pump a nd Acce ssorie s -
Ala rms and Safety Feat ures
Line Empt y
Re me dy
Ba t t e rie s a re not cha rging prope rly
Re me dy
Ba t t e ry is low
Re me dy
Ba t t e ry is de ple t e d
Re me dy
Freephone 24 hour Advice Line
The bag is empty.
The set is not fully primed.
The air sensor area is contaminated.
The infusion will stop.
Stop the infusion/ Change/ Check the bag and Ambix giving set.
Prime the Ambix giving set.
Clean (see cleaning procedures).
The pump is plugged in at the mains.
Contact Calea Patient Services or the helpline.
From the first alarm you will have 30 minutes of battery power left.
Plug the pump into the mains.
No external power available.
The infusion will stop.
Check the installation in the holder.
Check the contact pins and the pump.
Check the cable and connect to the mains.
Plug into mains to complete or stop infusion.
12
0 8 0 0 121 8 30 0
Use r Guide