Alcolizer Druglizer LE5 Manual do utilizador - Página 4
Procurar online ou descarregar pdf Manual do utilizador para Equipamento de teste Alcolizer Druglizer LE5. Alcolizer Druglizer LE5 19 páginas. Charging cradle
Também para Alcolizer Druglizer LE5: Manual do utilizador (16 páginas), Manual do utilizador (20 páginas), Instruções de montagem (3 páginas)
2 TEST PRINCIPLE
The Druglizer LE5 Drug Testing System is based on a competitive lateral flow immunoassay
technique where drug molecules that may be present in the oral fluid sample compete with
the same drug attached to the nitrocellulose wicking membrane, for limited binding sites on
tagged drug specific antibodies.
The antibodies are tagged with gold colloids to produce a conjugate with a red colour.
During testing, an oral fluid sample is absorbed onto the oral fluid collector pad. Chemicals
react with the saliva to remove sample interferences and a portion of this sample wicks
onto the conjugate pad where drugs in the sample react with tagged antibody specific for
that drug. The sample continues to wick by capillary forces along the nitrocellulose
membrane, onto which is attached the same drug of interest. If no drug is present in the
oral fluid sample the tagged antibody can bind to the drug bound to the nitrocellulose and a
red line is produced.
If drugs are present in the oral fluid sample at concentrations greater than the cut-off
concentration, the tagged antibody sites will be saturated with the drug and will not be able
to bind to the nitrocellulose bound drug, hence no result is produced. As the drug
concentration in the sample increases the colour intensity of the test line decreases. The
test strip is not intended to be viewed visually rather it is intended to be interrogated by a
meter. The instrument uses a sophisticated algorithm the determine the result.
3 MATERIALS PROVIDED
The Druglizer LE5 is supplied with the following:
•
Druglizer LE5 main module.
•
Druglizer LE5 drug test module.
•
USB cable.
•
Battery Chargers (240VAC and 12VDC vehicle).
•
Silicon protective cover, cartridge plug and wrist strap.
•
Soft shoulder bag or hard plastic carry case (optional).
•
Bluetooth printer (optional).
Updated user Manual and AlcoCONNECT Toolbox can be downloaded from
https://www.alcolizer.com/software-updates/
AlcoCONNECT Toolbox Version 1.6 or above is used to download test logs and configure the
Druglizer LE5. (For AlcoCONNECT Toolbox instructions please refer to AlcoCONNECT Toolbox
User Manual).
Cartridges are not included with the purchase of an instrument. Cartridges are sold
separately in cartons of 25 tests.
REF: Druglizer LE5 Drug Tester User Manual V7.docx
Uncontrolled when printed
Page 4
© Alcolizer Technology