12 SERVICE LIFE
12.1 Calibration alerts
The Druglizer LE5 requires calibration every six months to maintain optimum performance.
The instrument will display 'Service Life Pending' warning message 30 days before
calibration is required.
The 'Service Life Pending' will be displayed each time the instrument is turned on until
calibration is performed.
Press the Test Key to cancel the message and continue testing.
When the 'Service Life Pending' message period has expired, 'Service Life Expired' message
is displayed, and the instrument will lock preventing further testing until the instrument is
calibrated.
At any time, you can review the service dates.
The user is not authorized to maintain or calibrate the instrument in any way with the
exception cleaning of the instrument's housing. Any other form of maintenance must not be
carried out by the user. Opening the instrument will render any warranty claims null and
void. The instrument is factory calibrated. The instrument must only be repaired or
calibrated by authorized, qualified personnel.
The instrument may be damaged by liquid ingress. Please ensure that water and other
liquids are prevented from penetrating the instrument. If any fluid enters the internal parts
of the instrument, switch off the instrument immediately and return to authorised service
centre.