Alcolock B-2 Series Manual de Calibração - Página 7
Procurar online ou descarregar pdf Manual de Calibração para Equipamento de segurança Alcolock B-2 Series. Alcolock B-2 Series 8 páginas. Alcohol gas standard
TROUBLESHOOTING
In the event of a calibration failure, first check the common causes of errors:
• Plastic tubing has been overused or has condensation present
• The alcohol gas standard value is not 130 PPM, or the value has been incorrectly entered in
the Select Value menu
• The alcohol gas standard is expired
• All connections to and from the gas valve are not secured properly
• The value entered in Altitude menu is incorrect
ALTITUDE REFERENCE TABLE
The altitude adjustment is in increments of 200 meters. Round the altitude up or down to the
closest value as follows:
CALIBRATION MANUAL
IF ALTITUDE IS
0 - 100 m
101 - 300 m
301 - 500 m
501 - 700 m
701 - 900 m
901 - 1100 m
1101 - 1300 m
1301 - 1500 m
THEN SET TO
0
200
400
600
800
1000
1200
1400
5