Alcolock L OEM Manual de Calibração - Página 5
Procurar online ou descarregar pdf Manual de Calibração para Equipamento de segurança Alcolock L OEM. Alcolock L OEM 8 páginas. Alcohol interlock, alcohol gas standard
NOTE: Press and hold the left button anytime to cancel time set and to return to the
Date / Time screen.
d. With the time adjusted, keep pressing Next until Set is displayed, and press Set.
e. Press OK. The SELECT TYPE menu is displayed.
8.
Choose Dry Gas and press OK. The SELECT UNITS menu is displayed.
8.
Select BrAC and press OK. The SELECT DIGITS menu is displayed.
9.
Select 2 digits and press OK. The SELECT VALUE menu is displayed.
Read the next section.
CALIBRATING A HANDSET (CONTINUED)
1.
The SELECT VALUE menu contains the gas standard concentration options. Verify your
canister label, select the appropriate PPM value, and press OK. This enters the Altitude
menu.
Altitude is preset to 0 (meters).
2.
Determine the altitude of your exact location.
NOTE:
• Find exact altitude online
• Altitude may vary within general region or city; always use exact value
3.
If your altitude is:
• 100 m or lower – leave at 0 m. Press OK.
• 101 m or higher – press [+] to adjust value (increases by increments of 200 m). Round
to the nearest increment (see table at back). Press Accept.
4.
A Wait message is displayed followed by a 2 minute warm-up. The handset is ready for
calibration when the TURN ON GAS message is displayed.
5.
Insert the delivery tube from the regulator valve into the mouthpiece port. Keep hold of the
handset as the delivery tube is rigid.
6.
Press the valve down to release the gas while keeping hold of the handset (see figure).
A Blowing message is displayed and the handset emits a tone.
CALIBRATION MANUAL
Hold handset
Press valve
3