Direct Healthcare Group Gate EASY SoftBack Manual técnico - Página 9
Procurar online ou descarregar pdf Manual técnico para Cadeira de rodas Direct Healthcare Group Gate EASY SoftBack. Direct Healthcare Group Gate EASY SoftBack 20 páginas.
5. Replacement of remote control
Incorrect remote control is easily changed by releasing the contact / remote control on the left pillar - new
remote control is mounted in the same way.
NOTE: It is important that contact is locked with the clamp lock to ensure IP X5.
6. Calibration of actuators
The calibration of the actuators (height / lowering only) is normally done fully automatically each time the chair
reaches its end positions (min height, max tilt).
If the chair still does not go in parallel or if the tilt function does not seem to work, an ACTIVE calibration can be
done. This is done by holding down <button-slow / up> + <button-slow / down> simultaneously.
After about 15 seconds the chair will be moved to its end positions ie. my height and maximum slope
backwards. Hold down the buttons until the end positions are reached. Release the buttons and wait 5
seconds. The chair is now calibrated and can be used again.
Calibration should NOT be done with user in chair. Note that all 3 actuators move during the calibration
phase. Otherwise, cancel the calibration immediately and contact Customer Service / GATE (or your
local dealer).
7. Reset service indicator
The remote control has a service indicator for periodic maintenance reminder (min once a year). The service
indicator measures voltage and is activated after 11 months (flashes every 4 seconds). After 12 months the
indicator will glow steadily.
The indicator is reset by holding down buttons on the remote control (top line, <fast up>, <fast down>) for 30
seconds.
TEC HN I CAL MANU AL
G ATE RD .SE
9