Alcovisor Jupiter Manual - Página 4

Procurar online ou descarregar pdf Manual para Equipamento de teste Alcovisor Jupiter. Alcovisor Jupiter 8 páginas.
Também para Alcovisor Jupiter: Manual de instruções (12 páginas), Manual de instruções (12 páginas)

Alcovisor Jupiter Manual
ii. Left boxes: the fields needed to be shown and printed
iii. Right boxes 1: the previous data inputted will be kept until the
user changes it.
iv. Right box 2: set compulsory entry.
v. Touching the name of the field to change it.
vi. A header of the printout can be set by entering the words in the
field "Header".
7. Alert: Allows user to choose a BAC point where result will display in
RED and beep. User selects the value (0.000 to 0.100%) and any
test result at or above this value will display in red.
8. Blanklock: The password to access the Blanklock option is 118118.
This optional lockout/timer will cause the Jupiter to begin a 15-minute
countdown timer when a breath sample with a result of 0.020 % or
higher is detected. To turn on this feature, enter Blanklock and select
Yes. To turn off this feature, enter Blanklock and select No.
9. Calibration: The password to access the calibration option is
119119. Please refer to the calibration instructions included in this
manual for proper calibration procedures.
10.
Calibration Period:
a. Select the number of days in the first column and number of tests in
the second column that may be allowed to occur before calibration
is needed.
b. Select "LOCK" to lock the device if calibration is needed.
c. Default is 360 days, 700 tests and unlocked.
PAS recommends all Jupiter units be calibrated annually or whenever an accuracy
test shows that BAC readings are no longer within tolerance limits. Because precise
calibration is crucial, it can be performed only by PAS Systems International or by
persons who have been properly trained for alcohol detector calibration. (Calibration
of the Jupiter must be performed using a NHTSA-approved wet bath simulator or
certified dry gas standard.) Successful completion of calibration should be recorded
in a calibration log.
Calibration