Pre-test Conditions
1. Operating temperature range is -10°C to 50°C. Use within this temperature
range will limit condensation from a subject breath to insure an accurate
reading from a breath sample.
2. It is recommended that the operator to observe a subject for about 20 minutes
before administering the breath test to insure that nothing is consumed or
expelled from the mouth of the subject.
eliminate reading from mouth alcohol.
Test
1.
Attach a disposable mouthpiece onto the sample port.
2.
Input information in touch screen.
NOTE: Maximum 16 characters for each input.
3.
When displays "Please Blow", take a deep breath and blow until BEEP tone
stops.
4.
Test result displays on the display.
5.
If printer is using, select "Print" to print the test result.
6.
Select "Retest" to start another test.
7.
If testee did not blow within 30 seconds, tester can select "Refuse" to confirm
the testee refused to do the test; or select "Test Again" to allow testee to do
the test again.
8.
If testee failed to give a valid breath sample, tester can select "Discontinue"
to confirm; or select "Test Again" to allow testee to do the test again.
9.
A passive test can be performed by select "Manual" during the testee is
blowing.
10. Push the mouthpiece removal button to remove the mouthpiece.
11. 16,000 test records can be stored, if memory limit is reached, new test is not
available. Send the device to service centre to upload test results to PC
and clear out all memories and use it again.