Calex PMB21 Manual do operador - Página 5

Procurar online ou descarregar pdf Manual do operador para Acessórios Calex PMB21. Calex PMB21 17 páginas.

Calex PMB21 Manual do operador
WIRING
Check the distance between the sensor and the Modbus Master. If necessary, the sensor can
be ordered with a longer cable attached.

POWER SUPPLY

Be sure to use a 24 V DC power supply. The maximum current draw is 50 mA per sensor.

MECHANICAL INSTALLATION

All sensors come with a 1 m cable and a mounting nut. The sensor can be mounted on
brackets or cut outs of your own design, or you can use the fixed and adjustable mounting
bracket accessories which are shown below.
Note: The sensor must be grounded at only one point, either the cable shield or the sensor
housing, but not both.

AIR PURGE COLLAR

The air purge collar below is used to keep dust, fumes, moisture, and other contaminants
away from the lens. It must be screwed in fully. Air flows into the 1/8" BSP fitting and out of the
front aperture. Air flow should be no more than 5 to 15 litres/min.
Clean or 'instrument' air is recommended.
5