ACONLABS On Call Vivid Inserir manual
Procurar online ou descarregar pdf Inserir manual para Medidor de glicose no sangue ACONLABS On Call Vivid. ACONLABS On Call Vivid 2 páginas.
REF G125-122
PRINCIPLE AND INTENDED USE
®
The On Call
Vivid Control Solution contains a known concentration of glucose. It is used to confirm that
®
your On Call
Vivid Test Strips and On Call
working together properly and that you are performing the test correctly.
You should perform a quality control test:
•
Before you use your meter for the first time. This will help you get used to this test.
•
Before using a new box of test strips.
•
When you suspect that the meter or test strips are not working properly.
•
When you suspect that your test results are inaccurate. Or if they do not match with how you feel.
•
If you suspect your meter is damaged.
•
At least once a week.
•
After cleaning your meter.
Two levels of control solution are available. They are Control Solution 1 and Control Solution 2. Control
Solution 1 is sufficient for most self-testing needs. If you think your meter or strips may not be working
correctly, you may also want to do a level 2 test.
Control Solution 1 contains less than 0.2% glucose (active ingredient) and Control Solution 2 contains
less than 0.4% glucose (active ingredient), both have preservatives in an aqueous based mixture.
STORAGE AND HANDLING
•
Store the control solution at 36-86°F (2-30°C).
•
Do not refrigerate or freeze.
•
If the control solution is cold, do not use until it has warmed to room temperature.
•
Use before the unopened expiration date that is shown on the bottle.
Note: All expiration dates are printed in Year-Month format. 2013/01 indicates January, 2013.
•
Use the control solution only for 6 months after you first open it. The control solution will expire 6
months after the bottle is opened for the first time. Record this opened expiration date on the bottle label.
•
For in vitro diagnostic use. Use the control solution to test only outside the body. Do not swallow or
inject. For self-testing use.
•
Shake well before using.
•
To get accurate results, do control solution testing between 50 and 104°F (10-40°C).
•
The control ranges shown on the test strip vial (or on the foil pouch) are not a recommended range
for your blood glucose level. Figure out your personal blood glucose target ranges with your doctor.
•
Do not touch the end of the test strip to the control solution bottle. This could cause contaminants to
enter the control solution bottle.
®
•
Use only On Call
Vivid brand control solution with your On Call
®
meter and On Call
Vivid Test Strips.
MATERIALS PROVIDED
•
Control Solution
Please contact Customer Support at 1-800-838-9502 for more information on obtaining a control
solution kit.
Blood Glucose Control Solution
Package Insert
English
®
®
Vivid or On Call
Vivid Pal Blood Glucose Meter are
COMPOSITION
PRECAUTIONS
®
Vivid or On Call
•
Package Insert
• Meter
1. Insert a new test strip to turn on the meter. Refer to your meter's User's Manual details on how to
record the result as a quality control test, and more details on operating the meter.
2. Shake the control solution bottle thoroughly.
3. Squeeze the control solution bottle gently. Discard the first drop. If the tip clogs, tap the tip gently on
a clean, hard surface. Shake again, and then use.
4. Squeeze out a second small drop on a clean nonabsorbent surface. Touch the sample tip of the test
strip to the control solution drop. Ensure the strip gets enough sample.
Notes: Do not apply control solution to the test strip straight from the bottle.
If the control solution sample does not completely fill the check window, do not add a second
drop. Discard the test strip and start over with a new test strip.
5. Read the result from the meter display.
Make sure the control solution test results are in the control range. The ranges for both CTRL 1 and
CTRL 2 are displayed on the test strip vial (or on the foil pouch). For confirmation of results, Control
Solution 1 tests should fall within the CTRL 1 range. Control Solution 2 tests should fall within the CTRL
2 range. If the test results are in the respective ranges, this means your Blood Glucose Monitoring
System is working right and you are doing the procedure correctly.
If the control solution tests results do not fall within the respective ranges:
•
Check the expiration date of the test strip and control solution. Make sure that the test strip vial and
the control solution bottle have not been open for more than 6 months. Throw away any expired test
strips or control solution.
•
Make sure you are testing at a temperature between 50 and 104°F (10-40°C).
•
Make sure that the test strip vial and the control solution bottle have been tightly capped.
•
Confirm that you are using On Call
•
Make sure that you followed the test procedure correctly.
After checking everything listed above, repeat the control solution test with a new test strip. If your
results still fall outside the range indicated on the test strip vial label or on the foil pouch, your meter may
not be working properly. DO NOT use the system to test blood. Contact Customer Support for help. For
complete instructions, please refer to the User's Manual included with your meter. For additional
questions or issues with this product, please contact Customer Support at 1-800-838-9502. Customer
Service is open 24 hours a day, 365 days a year.
®
Vivid Pal brand
MATERIALS REQUIRED BUT NOT PROVIDED
•
Test Strips
INSTRUCTIONS FOR USE
EXPECTED RESULTS
®
Vivid brand control solution.
ACON Laboratories, Inc.
10125 Mesa Rim Road
San Diego, CA 92121, USA
Customer Support 1-800-838-9502
www.acondiabetescare.com
Number: 1150645402
Effective Date: 2013-01-11