Symptom
Possible Cause(s)
No power/green
1
Bed is not plugged into AC power.
power indicating
2
Power cord is cut/damaged.
light not on
3
Loose plug for handset.
Cut/damaged handset cord.
4
Actuator plug not connected.
1
Actuators operate,
Cut/damaged handset cord.
but the wrong
2
Damaged/defective Y-cable.
movement occurs
One or more
1
Loose or unplugged connection
actuators do
of actuator cord to control box.
not function
2
Cut/damaged actuator cord.
3
Defective actuator.
4
Defective control box.
5
Damaged/defective Y-cable.
Bed is not level
1
Actuators are out
when operating
of synchronization.
Hi-Low functions
2
Damaged/defective Y-cable.

troubleshooting guide

Action
1
Check plug to receptacle.
2
Check power cord for damage.
3
Have qualified technician check
115V outlet.
1
Inspect handset cord for damage.
2
Unplug the Y-cable and plug
the handset directly into the
control box.
1
Check actuator plug to control box.
2
Inspect actuator cord for damage.
3
Replace actuator.
4
Replace control box.
5
Unplug the Y-cable and plug the
handset directly into the control
box port.
1
Run bed to full down position
and continue to hold button for
3 seconds (after bed reaches full
down position).
2
Unplug the Y-cable and plug the
handset directly into the control
box port.
1-800-634-7328
DirectSupply.com
7