9. Installation
This microscope is a precision instrument. Using the microscope in an unfavorable environment could
result in malfunctions or degraded performance. Consider the following conditions when choosing the
installation location.
• Observation conditions are better if light from windows and bright room light can be avoided.
• Install the microscope in a location with a room temperature of 0˚ to 40˚ C and with a maximum
relative humidity of 85%. High temperature and humidity are to be avoided because they promote
mold growth and condensation, which may damage the microscope.
• Dirt and dust degrade optical performance and are to be avoided.
• Vibrations in the environment will degrade the image. Install the microscope in a location free of
vibrations.
• Install the microscope on a solid table and keep the microscope level.
• This microscope emits a feeble electromagnetic wave. Do not place a precision electronic device
near the microscope as precision could be degraded. Also, avoid placing a radio or TV near the
microscope as reception of sound and images may be hampered.
• Install the microscope in the location of 10 cm or more away from the surrounding walls.
• Install the microscope in the place that allows easy detachment of the power cord from the AC inlet
in the event of emergency.
10. Carrying the Microscope
This microscope is a precision instrument. Handle it gently.
Strong shocks and forcible operation will damage the
instrument. Shocks to the objectives, especially, could degrade
image precision.
• When carrying the microscope, hold it at its arm with both
hands.
• Do not hold the focus knobs, the eyepiece tube, or the
stage. These parts could easily come off and could result
in malfunctions.
11. Handling the Lamp
Do not touch the glass part of the lamp with bare hands. Wear gloves or use a cloth when handling the
lamp so as not to leave fingerprints on the surface. Wipe off any fingerprints or stains using a clean
cloth moistened with alcohol. Fingerprints will etch into the hot surface of the lamp and reduce the
brightness, damage the lamp or reduce its service life.
Handle the lamp gently. Shocks and vibrations will damage the lamp or reduce its service life.
When changing the lamp, be sure that the contact is not damaged. If the contact is damaged, the lamp
may not light up or may overheat. Insert the lamp's contact pins fully into the socket holes. If the pins
are loose, the lamp could come off or result in a contact failure, which will cause overheating or smoke.
Also, make sure that the field lens unit is securely attached.