Nikon NR-5500 Manual de reparação - Página 8

Procurar online ou descarregar pdf Manual de reparação para Instrumentos de medição Nikon NR-5500. Nikon NR-5500 25 páginas. Auto refractometer

Nikon NR-5500 Manual de reparação
3 . Refractometer Test Items
TEST PROGRAM
Check chopper frequency
and R/L. FOG and AX sensors.
Turn on M. LED.
Rotate IRM.
Not used.
Adjust alignment mark.
Turn on C. LED.
Turn on M. LED.
Not used.
Measure count offset value.
Measure CHA-FIL. CHB-FIL
peak values.
Not used.
Check monitor background
color.
Test printer operation.
Adjust automatic fogging
target position.
Adjust key display position.
Making of model eye count
table.
Adjust alignment offset.
Check a measurement system.
Initialize memory.
Check touch panel switches.
Adjust cylindrical axis
offset.
Check 1C card reader/writer.
Selection model eye mode and
human eye mode.
CHOP=524 ±1.0
Light occlusion = LO
MLED ON: M. LED lights up
continuously.
IRM ROT: Rotates continuously.
TOT=80—85: Dust-proof glass is
instal led.
TOT=90—95: Dust-proof glass is
not installed.
CLED ON: C. LED lights up
continuously.
MLED ON: M. LED lights up
continuously.
PHASE COUNT OFFSET=-200~+200
CHA=CHB=150 —200
0 V=000
Measurement switch
[PRINT] key
+: PRINT switch
-: Joy stickCmeasurement button)
Use the model eye set
(-18D- +22D).
Set to T = R±10.
Set to X = R± 10.
Set to Y = R±10.
R=1 2 0-1 3 5
A = B = 1 5 0 - 2 0 0
0
=
000
[X] [Y] = 60 when checking
[X] [Y] = 30 when adjusting
When EEPROM is replaced.
Verify that the switches can be
turned on and off normally.
Use J38056 cylindrical model
eye.
Valid only when an 1C card
reader/writer is built-in.
HUMN EYE = Human eye
MODL EYE = Model eye
+ 1.00-1.00 Dpt.
VR3 GAIN
VR6 TOT
VR5 XAL
VR4 YAL
VR7 CAL
C6- 2]
Repair Manual