Replacing the Battery (continued)
2
Install the battery with the positive (+) sign up
towards you.
3
To close, slide the battery door in until it snaps
into place.
Note: The European Battery Directive requires
separate collection of spent batteries, aiming to
facilitate recycling and to protect the environment.
The batteries in this product should be removed and
disposed of in accordance with local regulations for
separate collection of spent batteries.

Caring for Your Meter

Avoid getting dirt, dust, blood, control solution, water or any other
liquid in the meter's test strip and data ports.
Important: Never immerse the meter in water or any other liquid. This
may cause inaccurate reading or meter malfunction.
Do not try to clean the strip or data ports.
Do not pour liquid into the strip port, data port or buttons.
Do not place your meter in water or any other bath.
Clean the outside of the meter using a cloth dampened with one of
the following:
• Mild detergent/mild soap and water.
• 70% isopropyl alcohol.
• A mix of one part household bleach, nine parts water.
27
www.AbbottDiabetesCare.com

System Troubleshooting

Error Messages

Error 1
Possible Causes
• Sample is too small.
• Problem with the FreeStyle Lite test strip.
• Problem with the meter.
• Very low blood glucose (less than 20 mg/dL or
1.1 mmol/L).
What You Should Do
1. If you have symptoms such as weakness, sweating, nervousness,
headache or confusion, follow your doctor's recommendation for
treating hypoglycaemia.
2. Conduct a control solution test using a new FreeStyle Lite test
strip. If the results of the control solution test are within the range
printed on the side of your test strip vial, retest using blood and a
new FreeStyle Lite test strip.
3. If the retest using control solution does not work or the error
persists, call Customer Care.
Error 2
Possible Causes
• Problem with the FreeStyle Lite test strip.
• Problem with the meter.
• Very high glucose (above 500 mg/dL or
27.8 mmol/L).
• HIGH control solution applied when
temperature is too cold. (This applies only to
control solution that is labeled as HIGH.)