Camille Bauer SINEAX CAM Manual de Instruções - Página 7
Procurar online ou descarregar pdf Manual de Instruções para Instrumentos de medição Camille Bauer SINEAX CAM. Camille Bauer SINEAX CAM 12 páginas. Graphic display for sineax cam
Também para Camille Bauer SINEAX CAM: Manual de Instruções (19 páginas)
Symbol
Meaning
Device may only be disposed of in a professional
manner!
Double insulation, device of protection class 2
CE conformity mark. The device fulfi lls the require-
ments of the applicable EC directives.
Caution! General hazard point. Read the operating
instructions.
General symbol: Input
General symbol: Output
General symbol: Power supply
6.1 Software installation
For the parametrization of the device the supplied PC software
CB-Manager has to be installed. To do so, execute the fi le
setup.exe in the CB-Manager directory on the CD.
The fi le "Read Me First.pdf" on the software CD
contains all information about the installation of
the CB-Manager software as well as assistance
for problems concerning USB support.
6.2 Parametrization
The usage of the softwarer is comprehensively described
under Help | Contents. There you will fi nd also all detail infor-
mation, which may also be requested context specifi c. Here-
inafter an overview about existing help topics is shown.
The device confi guration may be performed ONLINE (with
existing connection to the device) or OFFLINE (without con-
nection to the device). Select Parameter | Edit in the device
menu to see an overview of the present settings. You always
will be asked if the present parametrization should be read
from the device.
The complete confi guration of the device is devided in subjects
and displayed in a register form. This type of display is well
Pos.
Meaning
1
Works No., Manufacturing date
2
Terminal connections relay outputs
3A
Terminal connections current inputs
3B
Terminal connections when using Rogowski coils
4
Terminal connections voltage inputs
5
Input frequency range
6
Terminal connections power supply
known from the control panel of Windows. In each register
you may request context specifi c help. Therefore this manual
will only describe some functions using elements from more
than one register.
Sequence
When entering the device confi guration you have to consider
a useful sequence. After defi ning the device hardware in the
register device you have to enter the input parameters, be-
cause all the following inputs will depend on these settings.
A help to do this is the function "Next", which navigates you
through the registers in a predefi ned sequence, depending
on the selected hardware. This function minimizes possible
dependencies.
State signalling to digital or relay outputs
Only via logic module it is possible to output a determined
logic state to a digital or relay output. Logic states may be
determined from limit value states, states of digital inputs,
values received via bus interface or previous calculated logic
states. But all these possible logic inputs have to be defi ned
in advance: Limit values in the register limit value and digital
inputs in the register I/Ox, which must have the functionality
digital input. Further information may be obtained from the
help of the logic module.
Meters
Analog or digital inputs may be used to build meters. The
defi nition of the measurands to summarize is done via the
I/O registers of the appropriate I/O modules. The register
meters will then display all possible meters and provides the
possibility to activate the tariff switching to build high and low
tariff meters. This meters list does not contain the 12 active
and reactive power meters of the standard fi tting.
6.3 Simulation / Measurement acquisition
The behaviour of the I/O modules may be simulated during
commissioning. By setting states or measurement values it
is possible to test if following circuits show the correct be-
haviour resp. if the SINEAX CAM responds to input devices
the right way.
7