3.2 Normal Operation

After initial training and under normal conditions the following procedure
should apply:
Open the Plexiglass cover
1.
Check that the chute is mounted and that the proper collecting device of de-capped
2.
plugs is in order.
Place a loaded carousel with the colour mark in line with the home position pin on the
3.
drive disk. Check that the first tube is located correctly in line with the de-capper head.
Close the cover
4.
After power on, press the start button.
5.
After automatic stop, open cover and replace the carousel for continuous operation.
6.
Close cover and press start.
7.
After finishing operation, close cover and power off the unit.
8.
On daily basis, remove the chute and de-capper head for cleaning according the
9.
Cleaning/Service instructions.
Figure 4: Chute attached on Pluggo (standard on right-side)
To remove the chute: Lift and pull it straight out from side of Pluggo