3. Verify that the Lumaview 500/400 version is the latest version listed in Appendix A. The version
number can be found by clicking on the Help dropdown menu and selecting About.
G. Objectives
1. The LS500 is shipped with the installed objective fully retracted. When used for the first time, it is
necessary to turn the focus knob a number of turns to reach the sample focal plane. Turning the focus
knob clockwise raises the objective and counter-clockwise lowers the objective. The safest procedure
for focusing is to raise the objective until it almost touches the sample and then lower the objective
until focus is achieved.
2. To change objectives, use the supplied 2 mm hex wrench to remove the 4 screws holding the back
panel in place. Move the panel away and with one hand reaching through the deck hole and the other
reaching from the back, carefully switch out the objective. Do not overtighten. Carry out the
replacement quickly to minimize the chance of dust falling onto the mirrored dichroic filter. If any dust
is observed on the filter, carefully blow it off using clean air such as from a compressed air can, making
sure to hold the can upright and bending the straw to aim the air directly toward the filter. If the can is
tilted, liquid can be emitted, which will require further cleaning. Other methods to remove dust may
damage the dichroic filter and must NOT be used.
3. For Meiji Plan Achromat and Plan Semi Apochromat objectives, the optional Objective Removal Tool
can also be used to change objectives. To use the Objective Removal Tool, lower it (grip end up)
through the center hole on the LS500 deck and match the inner notches of the Tool with the notched
pattern on the installed Meiji objective body. Unscrew the objective, and then lower and rotate the
Removal Tool just slightly so the notches no longer line up. This will allow lifting of the objective out of
the LS500 with the Tool. Reverse steps to install the replacement objective.
H. Power
1. For almost all imaging applications, power through the USB cable to the LS500 will be sufficient. If high
levels of illumination are required, e.g., with very dim samples, connect the electrical plug of the
included external power supply/cord into an electrical outlet and the small round plug into the round
port on the back of the LS500 (see Fig. 1).
The external power supply/cord with country-specific plug LS500 is for use with fluorescence and thus
the LS400 does not come with this power supply/cord.
I. Live Cell Imaging with LS500 in an Incubator
1. Prior to imaging in an incubator, place the LS500 inside for 12 hr to allow equilibration to the incubator
temperature.
2. Place the sample on the LS500, focus, and close the incubator door. After 60 minutes, check the focus.
If satisfactory, imaging can be initiated.
3. When live cell imaging at magnifications of 40x and above, the focus can sink several microns as the
mechanism relaxes. If the image has drifted out of focus, estimate the depth of focus change required
to bring the image back into focus and overcompensate the same amount past the point of focus.
Check the focus again after 60 minutes. The image should be in focus and imaging can be initiated.
(Repeat process as needed).