STERILIZATION:
WARNING -
between each patient use.
WARNING -
method or temperatures other than what
are prescribed may damage the motor
or present a risk of cross-contamination
between patients.
CAUTION -
Do not soak or submerge
the motor in any liquid.
STERILIZATION PROCEDURE:
Pre-clean
1) Thoroughly brush off any visible signs of
debris from the motor and cord.
2) Thoroughly clean the device with a moist
cloth or towel to remove any remaining signs
of debris.
Sterilize
3) Select one of the three following
sterilization methods (A. B. or C.):
Wrapped Sterilization –
appropriately sized sterilization pouch and
seal it.
A. Gravity Wrapped:
• Minimum Temperature:
132° C (270° F)
• Full Cycle Time: 20 minutes
• Minimum Dry time: 30 minutes
B. Prevacuum Wrapped:
• Preconditioning Pulses: 3
• Minimum Temperature:
132° C (270° F)
• Full Cycle Time: 4 minutes
• Minimum Dry time: 40 minutes
Flash Sterilization – For immediate use
only.
C. Gravity Unwrapped:
• Minimum Temperature:
132° C (270° F)
• Full Cycle Time: 10 minutes
No dry time is required for flash
sterilization.
MOTOR & CORD ASSEMBLY:
•
The entire motor and cord assembly is fully
autoclavable. Before autoclaving the
assembly, remove any handpiece from motor
6
Sterilize the motor
Use of a sterilization
Place in an
and install autoclaving plugs into the end of
the motor and motor cord connector as
shown below. The plugs are supplied with the
motor and cord.
• Loosely coil the motor cord when auto-
claving. Avoid sharply bending the cord when
autoclaving.
• Always install the autoclaving plugs into motor
and motor connector before autoclaving.
Turn black connector ring only
• If the autoclaving plug for the motor becomes
difficult to remove or install, apply silicone grease
or similar lubricant to the O-rings.
NOTE: Call Aseptico, Inc. at 1-800-426-5913
for any questions or clarifications on this
sterilization procedure.
MAINTENANCE & CLEANING:
HANDPIECES -
Thorough cleaning and
lubrication of handpieces after each use and
before sterilization is very important to ensure
proper operation and service life of the
handpiece. Follow the instructions provided with
the handpiece purchased for complete
maintenance instructions.
WARNING
• Do not attempt to disassemble the motor or motor connector.
• Do not oil or lubricate the motor.
• Do not attach a handpiece to the motor while motor is running.
• Do not bend motor cord sharply.
Failure to comply with any of the above instructions may
void your warranty
CONSOLE MAINTENANCE -
the console may be cleaned by wiping with a
soft cloth moistened with mild detergent or
disinfecting solution.
IRRIGATION TUBING SETS - Irrigation tubing
sets can be removed from the pump and may
be sterilized by autoclave, chemclave, or
ethylene oxide gas processes. Follow the
manufacturer's recommended procedure.
NEVER store the tubing sets in the pump for
extended periods of time. The tubing may
collapse and stick to itself creating a blockage
for future uses.
The exterior of