Water Off
Apply pressure to
activate Handpiece
Figure 8 - Foot Control
4.
WATER SUPPLY TANK - The Aseptico ADU-20B
incorporates a self contained pressurized water system (Fig.
9). This consists of a 2-liter HDPE tank dispensing water
through the 3-way syringe and handpiece control. The water
tank is pressurized to a constant 35 pounds per square-inch
gauge (PSIG) during use.
Before removing the tank for refilling, it must be
depressurized by turning the tank 1/4 turn counterclockwise.
The tank is removed by unscrewing it from its mount at the
bottom right rear of the chassis After the tank is filled, screw
it back onto the mount. Cleaning and sanitization of the
tank is recommended between uses.
Turn water supply tank 1/4
turn counterclockwise to
release pressure before
Figure 9 - Water Tank Removal
5.
PRESSURE GAUGE (Fig. 10) - The pressure gauge on the
right panel gives you a visual indication of the handpiece
drive pressure.
6.
WATER COOLANT FLOW CONTROL (FIG. 10) - Adjusts
the flow of water coolant to the handpiece control.
7.
FLUSH TOGGLE (FIG. 10) - The flush toggle located
on the right side of the chassis allows you to quickly
and completely flush your handpieces, washing away
contaminants which may have accumulated in the
handpiece and tubing. You should flush the handpieces for
about 5 seconds after every patient, and about 20 seconds
at the beginning of each day to reduce overnight bacterial
accumulation which may have occurred.
To flush your handpieces, remove them from their holder,
directing the spray away from you and into a basin, then flip
the flush toggle and hold the desired number of seconds.
Release the flush toggle.
8.
AA-21 HIGH VOLUME EVACUATOR ASSEMBLY - The
HVE assembly is an air powered, self contained oral
evacuator system. It is supplied with a plastic waste
container and a single high velocity hose.(Fig. 11)
Remove the vacuum bottle and hose assembly from
packing and install into the quick disconnect at the back of
the delivery head.
To operate, remove the vacuum hose from holder and insert
Water On
unscrewing
a standard oral evacuator tip. Depress the on/off switch on
the side of the hose end for vacuum.
Figure 10 - Right Side of Chassis
Water Coolant
Flow Control
Flush Toggle
Saliva
Optional
Ejector On/
Cavitron
Off
Connection
Waste from the vacuum system will collect in the attached
plastic bottle. When the waste container is 3/4 full it should
be emptied. After each use, the vacuum system should
be cleaned by running clean water or a vacuum system
cleanser through the evacuator hose and thoroughly
cleaning waste bottle. CAUTION: Use only NON-foaming
cleansers in the vacuum lines. A small and large capacity
waste container are available and are interchangeable. The
ADU-20B is supplied with a large capacity container.
Disconnect
Vacuum
On/Off
Figure 11 - HVE Vacuum System
9.
SALIVA EJECTOR ASSEMBLY - The Saliva Ejector is
an air powered, self contained oral evacuation system. It
is supplied with a plastic waste container and a single low
velocity hose (Fig.12).
To operate, remove the saliva ejector valve from the holder
and insert a standard saliva ejector tip. Turn the vacuum
toggle switch, located on the right side of the cart, to the
'ON' position. The lever valve assembly snaps apart at the
swivel and may be autoclaved.
Waste from the vacuum system will collect in the attached
plastic bottle. When the waste container is 3/4 full, it must be
emptied to prevent a back-up in the system. After each use,
the vacuum system should be cleaned by running clean
water or vacuum system cleanser through the evacuator
hose and thoroughly cleaning the waste bottle. CAUTION:
Use only NON-foaming cleansers in the vaccum lines.
Handpiece
HVE Connection
Pressure Gauge
Air Supply
Pressure
Regulator
Water Supply
Quick
Waste
Container
Air Supply
Line
Tank