Ashcroft 2132 Manual de Instruções - Página 11

Procurar online ou descarregar pdf Manual de Instruções para Instrumentos de medição Ashcroft 2132. Ashcroft 2132 20 páginas. Digital sanitary gauge

Ashcroft 2132 Manual de Instruções
Note: Calibration of Zero, mid-scale or span can be set inde-
pendently of each other. For instance, if only full-scale cali-
bration is required, press the
gauge display indicates full-scale pressure. Press Enter and
proceed as indicated above. Calibration of zero, midscale and
full-scale is recommended when recalibrating the gauge.
ZERO KEY (0bUTN): This feature allows the user to select
percent of full-scale at which the gauge can be rezeroed
using the Zero/Clear key on the keypad. Options include 5%
full-scale*, 10% full-scale or DISAB (disable of the zero key).
To use ZERO option:
Step 1: Press the Menu key on the keypad.
Step 2: Press the
key) until the word CONFIG appears.
Step 3: Press Enter.
Step 4: Enter user password, if it has been programmed.
Step 5: Press the
key) until the word 0bUTN (zero key) appears.
Step 6: Press Enter.
Step 7: Press the
key) to select 5PCT (5% full-scale), dISAb (disable zero
key) or 10PCT (10% full-scale).
Step 8: Press Enter to finalize the selection.
DISAB: or disable: This feature allows the user to dISAb
(or disable) or ENAb (enable) items in the MENU. Some
keypad keys can also be enabled or disabled. Any or all
MENU items can be enabled or disabled.
To use DISAB option:
Step 1: Press the Menu key on the keypad.
Step 2: Press the
key) until the word dISAb appears.
*Indicates default.
MENU FUNCTIONS
(down arrow key) until the
(up arrow key) or
(up arrow key) or
(up arrow key) or
(up arrow key) or
– 11 –
(down arrow
(down arrow
(down arrow
(down arrow