Installation and Maintenance Instructions for
F5509/F6509 Differential Pressure Gauges
Release the pressure on the axle, and remove
the adjustment key.
7.4.3 Contact function
Function 1: Close contacts with the indication rising in a
clockwise direction.
Function 2: Open contacts with the indication rising in a clock-
wise direction.
Contact assignment:
1st contact left target value indicator,
2nd contact middle target value indicator,
3rd contact right target value indicator
7.5 Subsequent relocation of the gauge (by
the customer)
Recommendation: Do not remove the pressure
gauge from one pressure monitoring location to another,
as there is a risk of the process media being mixed, with
unforeseeable chemical reactions.
8 SERVICING
The device is maintenance-free. However, to ensure reliable
operation and a long lifetime for the device, we recommend
that it is checked regularly.
8.1 Safety
When servicing the pressure gauge, the process lines must be
depressurized. Special attention should be taken to ensure
process pressure is not applied.
8.2 Gauge performance and calibration
Gauge performance and recalibration should be checked at
regular intervals, depending on the application. The precise
testing cycles should be adjusted in line with the operating
conditions and ambient conditions. In the event of various
device components interacting, the operating instructions for
all other devices should also be taken into account.
• Check on display.
• Check on function, in conjunction with downstream components.
• Check of pressurized connection pipes for seal condition.
• Check of electrical connections
8.3 Cleaning and maintenance
Cleaning is carried out using a non-aggressive cleaning agent,
with the ventilation valve closed and respecting the protection
category of the device.
WHEN ACTION IS REQUIRED TO REPAIR/REPLACE
9
THE GAUGE
9.1 Safety
Defective or faulty pressure gauges put the operational safety
and process safety of the plant at risk, and can lead to a risk or
injury to persons, the environment or the plant.
© Ashcroft Inc. 2018, 250 East Main Street, Stratford, CT 06614-5145, USA, Tel: 203-378-8281, Fax: 203-385-0357, www.ashcroft.com
All sales subject to standard terms and conditions of sale. F5509_gauge_im1.0, Rev. A, 08/18
9.2 Conduct in the event of failure
All defective or faulty gauges must be taken out of service. If a
repair is required contact inside sales.
9.3 When action is required to repair/replace the gauge:
Possible situations when action should be taken:
• Jerky or random movement of the pointer
• Pointer does not set to zero for pressure less display
• Indications that the measurement system seal (diaphragm)
has been breached (discoloration to dial display or of filling
liquid)
• Bent or loose pointer
• Cracked window
• Leaks when the device is filled
• Damage to housing
• Indications that the measurement system seal is imperfect
(discoloration to dial display or of filling liquid)
In these instances, repair or replacement of the pressure gauge
is always required.
9.4 Gauge repair/replacement
See Chapter 7.3 Mounting/Installation
10 REMOVAL, DISPOSAL
10.1 Safety
Residues of measuring media in and on removed
gauges can constitute a risk to people, the
environment and equipment. Adequate precautionary measures
must be adopted. If necessary, the devices must be cleaned
thoroughly (see advice in safety data sheets).
10.2 Removal
• When servicing the gauge, the pressure lines must be
depressurized, the electrical connections isolated from the
mains supply, and the plant secured against being switched
on again.
• Remove the gauge using a suitable tool
10.3 Disposal
Please help to protect the environment
and dispose of or recycle the devices and
components used in accordance with the applicable
regulations.
11 APPENDIX
11.1 Data sheet for Bourdon tube pressure gauge F5509/
F6509
Detailed data sheet is available from supplier's website (see 1.6
Manufacturer's address, customer services)
Model Description
F5509/
This table refers to specific documents:
SS differential pressure gauge F5509/F6509
F6509
K5500
Electrical contact devices for pressure & temp gauges.G1.K5500
Document
G1.F5509
5