Ashcroft 1305DMh-150 Manual de instalação e manutenção - Página 10
Procurar online ou descarregar pdf Manual de instalação e manutenção para Equipamento de teste Ashcroft 1305DMh-150. Ashcroft 1305DMh-150 20 páginas. Deadweight tester and portable pump
To replace the tester in its case, remove the gauge offset
pipe assembly (64 or 66) and the piston and cylinder
assembly (67 or 79). Install shipping plugs in body outlet
holes. Screw in the displacement valve (41) until it stops.
Close the vent plug (35). Open the release valve (37) ap-
proximately _ turn. Replace the tester in its case, reservoir
last. Engage the spring clip catch.
3.3.7 General Precautions
It is important that the deadweight tester be connected to
a leak tight system.
The deadweight tester should be set-up so that the axis
of the located piston is vertical, the weights carefully cen-
tered on that axis, and the piston rotated during use. The
purpose of the rotation is to spread the lubricant over the
entire surface between piston and cylinder, so that there
will be no metal-to-metal contact. If rotation is not main-
tained, the lubricant film will not cover the surface properly,
and readings will be in error.
The high and low pressure piston assemblies have an
internal overload stop, which prevents the piston and cyl-
inder from being forced apart if weights are accidentally
removed.
3.3.8 Operating Fluids
Standard Tester:
Any medium weight oil may be used (including automotive
oils S.A.E. 10, 20 or 30) S.A.E. 20-W recommended.
Hydraulic Tester:
Any hydraulic fluid (silicate or phosphate base). Skydrol or
Pydraul is suitable.
For normal operation, it is not necessary to change the
weight of oil for various pressures. A lighter oil may be
used where low pressures are being checked, and a
heavier oil where higher pressures are being tested.
The reservoir can be refilled while the tester is operating
under pressure.
3.3.9 Caution
Standard testers designed for oil service may not be used
with water for oxygen service.
3.3.10
Hydraulic service testers should not be filled with water or
any oil other than hydraulic fluid. Serious pump failure may
occur due to O-ring damage.
3.4. Installation Type 1327
3.4.1 Set up Procedure (See Figures 2-2)
To remove the test pump from the case, release the spring
clip, depressing the clip by hand, grasp the pump reser-
voir, and lift up the back end of the pump assembly.
Remove the two plastic shipping plugs from the outlet
connections. Remove the threaded filler plug (35) from
reservoir filling hole. Remove the offset pipe (64) or (66)
from the carrying case and connect to the inner pump
body outlet as shown in Figure 2-2. Remove the straight
pipe extension (58), and assemble to outer vertical pump
body outlet as shown in Figure 2-2.
3.4.2 Connecting The Gauge (See Figure 2-2)
Select a test gauge that is adequate for the pressure
range desired. Assemble the test gauge to the straight
pipe extension, using adapter nut (62), collar (61), adapter
(60) and reducer bushing (59).
Seal the connection by tightening nut adapter (62) and
adapter (60) until the coned end of the pipe is forced into
the gauge socket. Rotate the gauge for viewing by loos-
ening nut adapter (62), setting the gauge to the desired
position, and re-tightening nut adapter.
Repeat the above procedure to connect the gauge being
tested to the offset pipe assembly (64) or (66).
3.4.3 Bench Space Required
Allow a space:
10 inch (2.5 cm) width
23 inch (5.8 cm) depth
22 inch (5.6 cm) height allows for testing gauges up to 8.5
inch (250 mm) size
3.4.4 Storage Space Required
10
⁄
inch (27 cm) width
1
2
22
1
⁄
in (57 cm) depth
2
10 inch (25 cm) height
Storage space dimensions are larger than actual size of
test pump carrying case by approximately 0.5 inch (1.3
cm), in order to facilitate ease of handling in storage.
3.5 Operation 1327D
3.5.1 Test Gauge Selection
The portable test pump is used for calibrating instruments
such as pressure gauges, pressure switches or other
pressure devices rated up to 10,000 psi (70,000kPa).
First, select the proper test gauge, depending upon the
pressure range desired. When priming the pump a test
gauge rated to 10,000 psi (70,000kPa) should be used.
3.5.2 Priming The Pump
The offset pipe (64 or 66 Figure 2-2) and straight pipe
extension (58) should already be secured in place and the
reservoir should contain the operating fluid (refer to instal-
lation Section 3.4).
Close the release valve (37) by turning it clockwise and
open the air vent by loosening the filler plug (35) a few
turns. The shuttle valve (30) should be pulled outward
from the pump body, the end of the knob approximately
even with the edge of the drip pan. In this position the high
volume displacement mode is selected and the pump is
self-priming. Operate the pump handle several times using
full strokes, until you see fluid appear at the top of the
straight pipe extension. Connect the test gauge to it (refer
to Installation Section 3.4 for proper gauge connection).
Operate the pump handle a few more times until fluid
is observed at the end of the offset pipe. Connect the
gauge or other device to be tested to the offset pipe.
For purposes of priming, the device must be designed
to withstand the full operating pressure of 10,000 psi
(70,000kPa). If a lower range device must be used, ex-
treme care must be exercised to avoid overpressuring
its pressure element. Seal the connection to the device
by tightening nut adapter (62) and adapter (60) until the
coned end of the pipe is forced into its inlet. If necessary,
rotate the device for viewing by loosening nut adapter (62),
set it to the proper position and retighten nut adapter.
Again, operate the pump handle a few more times. After
a few strokes positive pressure will develop and the pump
handle will begin to resist pumping action. Raise the pump
handle to its uppermost travel position. Loosen the bleed
screw (51) a half turn counter clockwise and slowly op-
erate the pump handle through a downward stroke until
fluid flows steadily from the bleed vent. Close the bleed
screw just prior to completing the downward handle mo-
tion. Repeat this action until no air bubbles are observed
in the fluid flow. (Note – the bleed screw must be closed
when the handle is being raised). Push the shuttle valve
in towards the pump body and continue pumping to the
desired test pressure. Open the release valve (37) to vent
10
of 20