ASI 2045 Manual de operação e serviço - Página 12

Procurar online ou descarregar pdf Manual de operação e serviço para Equipamento dentário ASI 2045. ASI 2045 20 páginas. Dental operative unit classic assistant’s cart

c
laSSic
2.3
Handpieces (if so equipped)
Refer to the handpiece manufacturer's instructions for the proper connection, use and
maintenance of the handpiece prior to attaching them to this system.
Connect the handpieces to the appropriate tubing connector. The handpieces are
automatically activated whenever they are removed from their holder. (Ensure that only one
handpiece is removed from the holder at a time to prevent multiple handpiece activation. If
so equipped, engage the handpiece lockout feature for additional safety.)
Handpiece holders can slide from side to side on the bar, they can
ratchet around the bar, and they can be rotated. For instrument
handpieces, a manual lockout lever shuts the holder off so the
instrument can be removed without having it activated by the rheostat.
To engage the lockout, lift the lever to a horizontal position
A
. To return the holder to normal operation, point the lever back down
Figure 2.3.-B
to the floor.
2.3.1 Coolant Operation – Adjusting Coolant Spray
To activate the coolant water for a handpiece flip the toggle switch either located on the front of the system or on the foot
control. To deactivate the coolant water, flip the toggle switch to the OFF position.
To adjust coolant spray, locate the adjustment knob on the unit on the side or front. Turn the toggle switch for coolant spray
to ON, then turn the adjustment knob (clockwise to decrease, or counter-clockwise to increase) until the desired level of spray
is achieved. Note: For multiple instruments that provide coolant, individual adjustments are provided and are located in order
from left to right.
2.4

Optional Fiber Optics (if so equipped)

To determine the style of fiber optics you will need for your handpiece, look at the back of your handpiece or the swivel
coupler and compare the connections to the following drawings. For bulb replacement please refer to Service section of
this manual.
2.4.1 Five Hole Fiber Optic Systems
These systems are an adaptation of the four hole handpiece connection, simply by adding a fifth hole for the
light to enter the back of the handpiece. The light travels from the bulb up through the swivel coupler and
through the handpiece via fiber optic glass rods.
2.4.2 Six Pin Fiber Optic Systems
These systems are similar to five hole systems; however, instead of having the small bulb in the connector
at the end of the tubing, the bulb is contained in the swivel coupler from the handpiece. The connector at
the end of the handpiece tubing has two electrical sockets that connect to the two small wires. The swivel
coupler has two electrical pins that drop into these sockets when it is attached.
Unless specifically requested, units purchased with optional Universal Fiber Optics are set to a six pin
configuration at the factory. To modify the handpiece tubing to work with a bulb, remove the ISO adapter from the handpiece
tubing and insert the optional bulb ordered separately.
12
a
dvanced
ASI - The leader in Advanced Systems Integration www.asimedical.net
1044/1045/2044/2045
Figure 2.3.-
Figure 2.3.-A
Place bulb
into slot.
m
anual
Figure 2.3.-B
First, remove
adapter.
Rev.A