Ambu 480500000 Руководство по эксплуатации - Страница 7

Просмотреть онлайн или скачать pdf Руководство по эксплуатации для Медицинское оборудование Ambu 480500000. Ambu 480500000 10 страниц. Ascope 5 broncho sampler set (part 2 of 2)

4. Use of BronchoSampler 60
Before using the BronchoSampler 60 for the first time, follow the steps described below�
Characters in grey circles

4.1. Preparation and inspection

Refer to the Instructions for Use part 1 of 2 for inspection, preparation and for the use of
the aScope™ 5 Broncho.
Preparation for use of the BronchoSampler 60 with aScope™ 5 Broncho.
1�
Carefully con nect the sampler bridge with the aScope™ 5 Broncho, by mounting the
sampler bridge on the aScope™ 5 Broncho suction connector 1 � Ensure a tight fit�
Push in the attachment lock to secure connection 2 �
2�
3�
Carefully connect the vacuum suction tubing to the suction connector according to the
supplier's manual 3A � If the vacuum suction tubing has a male connector, use the suction
adapter to ensure fit 3B �
4�
Attach the sample container to the sampler bridge by joining the connection interfaces
and pressing the sample container towards the sampler bridge� Ensure the sample
container is clicked on 4 � Do not use preservatives in the sample container�
5�
To guard against potentially infectious materials during the procedure, wear personal
protective equipment�
4.2. Operating the BronchoSampler 60
1�
If suction is deemed relevant before sampling, ensure that the flow switch is aligned
vertically and pointing upwards 5 � Perform suction by pressing the suction button on
the aScope™ 5 Broncho in accordance with Instructions for Use part 1 of 2�
2�
To prepare for sampling of specimen, turn the flow switch counterclockwise towards the
sample container, and align it horizontally 6 �
Instillation of fluids should be performed in accordance with Instructions for Use part 1 of 2
3�
for aScope™ 5 Broncho�
4�
Perform sampling by pressing the suction button on the aScope™ 5 Broncho� Ensure that
the endoscope is held upright and level while sampling� Do not fill the sample container
above the maximum of the volume indicator scale�
5�
If suction is deemed relevant, turn the flow switch clockwise to align it vertically and
pointing upwards, and press the suction button on the aScope™ 5 Broncho� When suction
is complete, turn the flow switch counterclockwise towards the sample container, and
align it horizontally�
6�
Remove the sample container by pressing the flip top and pulling the sample container
from the sampler bridge 7 � Ensure that the endoscope is held level and that the flip top
is depressed during the removal of a sample container which holds fluid�
7�
Close the sample container by pressing the sample container lid towards the sample
container interface and engage the lid until it closes� Once the click-lock is engaged, do
not re-open the lid 8 �

4.3. After use

1�
If deemed relevant, detach the sampler bridge from the aScope™ 5 Broncho by
disengaging the attachment lock and pulling the sampler bridge off the aScope™ 5
Broncho� Disengage suction by pulling the suction tube off the sampler bridge�
2�
If deemed relevant, access the sample container by turning the sample container vial
clockwise 9 � It is recommended to handle contaminated substances in accordance with
local guidelines�
3�
Transportation and handling of a used sample container must be performed in
accordance with local guidelines for transportation and handling of hazardous material�
refer to the illustrations in the Quick Guide on page 2�
7