Co2meter CM-0155 Руководство пользователя - Страница 11

Просмотреть онлайн или скачать pdf Руководство пользователя для Регистраторы данных Co2meter CM-0155. Co2meter CM-0155 16 страниц. Co2 sampling data logger nema4

Sampling Data Logger User Manual
To ensure maximum accuracy, it is recommended to install the device in an environment that will 
routinely see this low value. Alternatively, the unit can be exposed to fresh air for a few minutes 
periodically, or set the ABC period to 0 sec. to perform a calibration as specified in the next section.  

Calibration

The calibration process varies depending on the type of unit and whether it has optional data logging 
functionality or not. All units are factory‐calibrated with multiple reference points of gas, and have been 
verified to be accurate within their specific functionality before shipment. However, if the unit is 
severely jolted or otherwise mechanically disturbed, the sensor can drift requiring recalibration. All 
calibration procedures follow a single‐point calibration routine that effectively shifts the zero‐point of 
the CO2 sensor. 
Attach calibration gas to the unit and connect it to a personal computer. Open the calibration screen in 
the GasLab® software. Click the "Calibrate" button in the calibration tab for the desired gas, located in 
the "Configure Sensor" screen. As long as the gas concentration is stable, the unit should instantly 
reflect the calibrated value. This can be confirmed by watching the display. To see the calibration value 
in real time, click in the "Collect Real time" button to capture these values before opening the 
configuration screen.  
0‐1% or 0‐30% Sampling Unit with Data Logging
Calibration can be performed using either 0% CO2 calibration gas (typically nitrogen, argon, etc.), or 
using a fresh source of air, assumed to be approximately 400ppm. 
Open the calibration screen in the GasLab® software. Click the "Calibrate" button in the calibration tab, 
located in the "Configure Sensor" screen. 
 
Figure 8: Real‐time capture 
CO2Meter
 
 
PAGE 11 of 16