MAINTENANCE

In order to replace the AeroCure's filters, lamps, and reset the filter display the front cover of
the device needs to be removed. This is performed by removing six Philips-head screws from
the front and top of the unit. The prefilter is mounted to the rear of the front cover and the
HEPA filter is mounted at the top of the unit. The filter reset button for the display screen is
located directly behind the screen on the inside of the unit, so you must remove the cover to
reset the filter. Once the cover is removed turn the unit on, there is a limit installed that will
only allow power to the filter life screen when the cover is removed ( If the motor or lamps
power on turn the unit off immediately and call customer support). Press and hold down the
small black button located at the far left (when looking at the screen from the inside), until the
screen flashes and resets back to filter life one hundred percent, it may take a few seconds for
the screen to reset.
CAUTION- Unplug the unit prior to removing the front cover. Failure to do so may result in
electrical shock, ultraviolet exposure, injury or death.
MAINTENANCE SCHEDULE
HEPA and Prefilter
UV Lamps
CAUTION: Use only genuine Aerobiotix® Aero Cure One ™ replacement filters and lamps to ensure
proper filter fitment and function of your unit. Failure to use specified products voids warranty.
Filter life display screen will indicate when filters need changed.
When filter life hits zero, both filters will need replaced.
Change sooner if significant filter soiling or reduced airflow.
Every 8000 hours or every 2 years.
Sooner if indicator lamp fails to illuminate
8