Amersham Pharmacia Biotech INV-907 Руководство по эксплуатации - Страница 2

Просмотреть онлайн или скачать pdf Руководство по эксплуатации для Блок управления Amersham Pharmacia Biotech INV-907. Amersham Pharmacia Biotech INV-907 9 страниц.

6 When used as an injection valve connect
the tubing as follows:
Port
Connect to
1
Column inlet
2
Sample loop or bottom tubing
of Superloop
3
Injection port
4
Waste
5
Waste
6
Sample loop or top tubing of
Superloop
7
Outlet of the pump
7 When used as an upflow/downflow valve,
connect the valve as indicated in the user
documentation of your ÄKTAdesign
chromatography system.

3 Operation

The valve is controlled from UNICORN.
When used as an injection valve use
the instruction
InjectionValve
System Control:Manual:Flowpath
be set in 3 positions:
LOAD
position 1 (home position)
INJECT
position 2
WASTE
position 3
If the valve has not been activated it is in
position 1.
2
in
. The valve can
LOAD, position 1
1
Column
2
Sample
3
syringe
4
Waste
INJECT, position 2
1
Column
2
3
4
WASTE, position 3
1
Column
2
3
4
7
Pump
6
5
Waste
7
Pump
6
5
Waste
7
Pump
6
5
Waste
Valve INV-907
Instructions 56-8101-01 Edition AC