Calibrate the DataMan 100
7
Adjust the placement of 
the calibration symbol 
until the alignment 
graphics turn green. 
8
Wait for the calibration process to finish. The Setup tool will indicate when 
the DataMan reader is calibrated:
18  DataMan 100 Verifier Quick Reference
Perform Code Quality Verification
In order to perform Direct Part Mark verification your reader must be 
calibrated as described in the previous section. To read symbols and 
view verification results, perform the following steps:
1. 
Start the DataMan Setup tool and connect to your reader.
2. 
Enable the 
Turn Code Quality On icon in the tool bar. 
Code Quality On icon
4. 
Select the 
Results Display task pane. 
5. 
Place the symbol under the reader and press the trigger. The symbol 
must be centered within the field of view, and the reader must 
be perpendicular to the surface containing the symbol and at the 
correct rotation.
DataMan 100 Verifier Quick Reference  19