Doran DS4500 Техническое руководство - Страница 14

Просмотреть онлайн или скачать pdf Техническое руководство для Весы Doran DS4500. Doran DS4500 16 страниц. Infant lactation scale

Doran DS4500 Техническое руководство

Section 6. Troubleshooting

General Problem Resolution

Weight reading will not repeat or
scale does not return to zero when
weight is removed.
Scale overloads before scale
capacity is reached.
Scale will not come to zero when the
ZERO button is pressed.
Weight readings are not correct.
Scale drifts off zero.
Scale shuts itself off or will not turn
on.
12
Problem
What to Do or Check
Person or object other than that being
weighed is touching the scale.
Verify that there is nothing caught in the
feet area or interfering with the feet.
Verify scale calibration is correct. If
problem persists, recalibrate the scale.
Make sure that the scale is stabilizing.
After pressing the ZERO button, the scale
should zero as soon as it becomes stable.
Verify the scale calibration with an
accurate test weight. If the readings are
not correct, recalibrate.
Be sure the platform is on a flat surface
and all four feet are touching the surface.
Check for air currents and/or vibration
around the scale.
Press the On/Off button to turn on the
indicator.
Check that new batteries are installed
properly in the battery tray.