DORC 1100 Руководство по эксплуатации - Страница 5

Просмотреть онлайн или скачать pdf Руководство по эксплуатации для Медицинское оборудование DORC 1100. DORC 1100 18 страниц. Micro diathermy system
Также для DORC 1100: Руководство по эксплуатации (19 страниц)

2.2

Operating instructions.

A.
Check if the electrical supply the unit needs is correct
(at the back-side).
B.
Plug the unit into the AC source outlet.
C.1. Assembling accessories :
The electrodes (1102-1111) are inserted into the handle assembly (1101) after unscrewing the
chuck cap approximately one-half turn counterclockwise. After the electrode is oriented in the
desired position, the cap is screwed down with very little finger pressure. Avoid over-
tightening the cap as this may cause the threads to become stripped.
The electrode handle with cable (1101) is then plugged into the connector marked Instrument on the
front panel. NEVER disconnect the cable by pulling on the cable itself.
The cable should be disconnected by grasping the connector between thumb and the fore-finger and
pulling straight out.
If the cable is not disconnected in this manner, damage to the cable will result.
NEVER TRY TO SEPARATE THE CABLE FROM THE HANDLE AS BOTH
CONSTITUTE AN INTEGRAL ASSEMBLY.
C.2. Coaxial Electrodes 1113
To replace a non-infusion type electrode 1113, carefully remove the threaded electrode assembly
from its handle and replace with a new electrode using the same procedure as with the chorioretinal
electrodes and the electrode handle with cable.
IMPORTANT : For sterilization instructions see section 3 .
30300300C
D.O.R.C. MICRO DIATHERMY SYSTEM
3/16