Cole Parmer 1200 Series Руководство пользователя - Страница 4
Просмотреть онлайн или скачать pdf Руководство пользователя для Измерительные приборы Cole Parmer 1200 Series. Cole Parmer 1200 Series 18 страниц.

• Has been subjected to severe transport stresses
Performance
Carry out performance checks with particular reference to wavelength and absorbance
accuracy to ensure that the spectrophotometer is working within its specification,
especially when making measurements of an important nature. Performance checks are
detailed in this manual.
Radio Interference
For compliance with the EMC standards referred to in the EC Declaration of Conformity,
it is necessary that only shielded cables supplied by Cole Parmer are used when
connecting the instrument to computers and accessories.
Introduction
The newly redesigned Cole Parmer 1200/1201 Spectrophotometers are single beam,
general purpose instruments designed to meet the needs of the conventional laboratory.
Cole Parmer 1200/1201 is ideal for various applications, such as: Clinical Chemistry,
Biochemistry, Petrochemistry, Environmental Protection, Food and Beverage Labs,
Water and Waste Water Labs, and other fields of quality control.
Featuring a digital display of photometric result, easy operation, and wavelength range
of 325 nm~1000 nm, Cole Parmer 1200/1201 are ideal for measurements in the visible
wavelength region of the electromagnetic spectrum.
Working Principle
The spectrophotometer consists of five parts:
1) Light Source (Halogen Lamp) to supply the light;
2) A Monochromator to isolate the wavelength of interest and eliminate the unwanted
second order radiation;
3) A Sample Compartment to accommodate the sample solution;
4) A Detector to receive and convert the transmitted light to an electrical signal;
5) A Digital Display to show absorbance or transmittance. Figure-1 illustrates the
parts.
relationship between these
100%T
0 A
Light Source
Monochromator
Sample Compartment
Detector
Digital Display
Figure-1
Block Diagram for the Spectrophotometer
4