Installation of Dahl VarioDock™
Installation must be carried out by a qualified and
manufacturers' instructions for use and assembly,
experienced technician/fitter.
which must be followed when installing the lock
In the area where the wheelchair is to be locked into
plate and spacer.
the docking system, the floor must be flat along the
C. Be aware that bolts with art.no. # 502800 for secur-
full length and width of the wheelchair. The length
ing the lock plate are custom made bolts, which
of the flat floor must be such that it is possible to
are only available in one length. It is very important
manoeuvre the wheelchair into the docking system
that the installer carefully adjusts the bolts to the
without problems. It is essential that there is no dif-
correct length, which fits the individual wheel-
ference in the height of the floor from one side to the
chair's approved adaptation kit with spacers, if
other at the location where the docking station is to
applicable. If the bolts are too short to fully engage
be installed. It is also important that the lock plate is
into the entire thread depth, they will not be strong
fitted straight/horizontally under the wheelchair. If the
enough to withstand the load during a collision. If
above is not carried out carefully, there is a risk that
the bolts are too long, the batteries or other wheel-
the wheelchair may not be manoeuvred properly into
chair components may be damaged. Therefore be
place in the docking station.
careful when shortening the bolts.
1. Delivery of VarioDock
TM
Warning! Never use bolts other than original Dahl
A. VarioDock
TM
is installed on a plate for transport
bolts # 502800 for fixing the lock plate. Standard bolts
when delivered to avoid damage during transport.
are not strong enough to withstand a collision.
Remove this plate and the 6 bolts before installing.
3. Measuring ground clearance
See also instructions 5 and 6 on page 17.
A. For correct determination of the wheelchair's
2. Installing the lock plate
ground clearance, it must be measured with the
A. Install the lock plate and 8 mm spacer under the
user seated in the wheelchair (pay attention to
wheelchair in accordance with the wheelchair man-
correct tyre pressure). Carefully drive the wheel-
ufacturer's instructions, which can be downloaded
chair into the VarioDock
from our website, as well as instructions on the
VarioDock
TM
drawings on pages 5 and 14. Before the lock plate
the underside of the lock plate and docking station.
and spacer are installed, carry out the following:
B. Read the ground clearance off the display and write
B. The installer must make sure that the wheelchair
it on the supplied sticker (see the design of the
has a solid construction at the place where the
sticker on the right).
lock plate shall be installed which can withstand
4. Location of VarioDock™
the load during a collision. On our website http://
A. Place the wheelchair, with the user sitting in it,
wtors.com you can see the list with all approved
facing forward in the vehicle at the desired installa-
adaptation kits for wheelchairs and download
tion location. If the user is the driver of the vehicle,
16
Maintenance manual, installation and user guide UK
make sure that the user can reach and operate all
necessary instruments and driving functions in a
proper and safe manner from this position.
B. Place the docking station under the wheelchair.
C. Mark the docking station's position on the floor
and the lock plate's position (in width and length)
under the wheelchair.
5. Beginning installation work
Ask the wheelchair user to leave the vehicle, so that
installation work can begin.
A. Drive the VarioDock
to the lowest position
TM
(61mm) before you begin installation work.
B. Loosen the 4 bolts on the side of the VarioDock
(important: note instructions on page 21!) and
carefully remove the top section of the docking sta-
tion. See pictures 3 and 4 on page 17. (Please note
that electronic components are installed in the top
section!). Mark the location of the bolts on the floor
of the vehicle, and drill holes through the floor
at these locations. Be careful when marking and
drilling through the vehicle floor to avoid damage
to brake pipes, wires, fuel tank and the like. Never
drill any holes until you are certain that the holes
Varenr. DIV99L - Labels, matsølv
Varenr. DIV99L - Labels, matsølv
match the location of the installation components.
and set the height of
TM
C. Secure the docking station with the supplied bolts,
so that there is 2 mm space between
Format:85 x 20 mm, udstanset
Format:85 x 20 mm, udstanset
nuts, etc. See illustrations pages 18 and 19.
Farver: Sort tryk
Farver: Sort tryk
Installation of Dahl VarioDock
1
TM
5
Removal of top section
Pictures 1-6
1. Loosen bolts, see instructions 3 on page 21
2. Bolts are loosened
3. Carefully remove the top section
4. Place top section as shown on picture
5. Plate and 6 bolts used for transportation
of the VarioDock
, shown on picture 6,
TM
must be dismantled and discarded!
6. The transportation plate and bolts are
now removed
TM
2
6
!
Fitting the top section
Picture 7
When assembling the slide and bottom
frames, the pin of the potentiometer must
be carefully placed in the hole of the bottom
frame, otherwise there is a risk of damaging
the potentiometer. If the docking station is
Drill holes in
tightened when the pin is not inserted in the
the anchorage
hole, the potentiometer will be destroyed/
plate for running
damaged.
through cables, if
necessary.
3
4
7