Baumer Hubner HENQ 1100 Руководство по эксплуатации - Страница 26

Просмотреть онлайн или скачать pdf Руководство по эксплуатации для Измерительные приборы Baumer Hubner HENQ 1100. Baumer Hubner HENQ 1100 36 страниц. Encoder analyzer

OPERATION / MENU DESCRIPTION
7.2.6

Error messages

)
When an error message appears, press the enter key to see the next message�
To disable error messages, set the value
Fig.2, page
Dead channel
Quadrature Error
Glitch detected
Missed Pulses
Wrong PULSE-No
Low Voltage
Sticky pin
Mark-Space
MaxSpeed
Phase
Missed ZeroPulse
ZeroPulse incorrect
Output Error
Low Battery Voltage
23/32
15�
Signal is missing
(for encoder with two channels displaced by 90°)
Illogical signals occurring / status
(for encoder with two channels displaced by 90°)
Glitch occurred
Missing pulses
Wrong number of pulses
U
or U
1
Did not reach signal level
Deviation of pulse/pause ratio is larger than
the set value (Default 30:70)
Maximum speed setting exceeded
Phase shift is larger than the set value
(Default ±18° of nominal value 90°)
Missing zero pulse (reference signal)
Incorrect reference signal
Output driver overloaded
Accumulator must be recharched
(only in devices equipped with accumulator)
Errors to off in the menu
too small
2
MB112EN - 11185046, 19A1, Baumer_HENQ1100_OI_EN
, see