i) Enter the measured resistance value according to the measuring record of the calibration
box in the text field.
j) Click on Calibrate.
k) Repeat steps i to k for all measuring points that are to be measured.
Connect the cable plug with the corresponding resistor.
l) Press button
Result:
ü
Calibration of measuring range completed.
INFO
This calibration method does not consider any change in the sensor geometry (e.g. by abra-
sion). If the sensor tip changes dimensions, the measured conductivity value will be different
from the previous one measured with the original tip geometry.
Any change in geometry can be compensated with the installation factor.