Doza DROZD Руководство пользователя - Страница 3
Просмотреть онлайн или скачать pdf Руководство пользователя для Измерительные приборы Doza DROZD. Doza DROZD 11 страниц. Dosimeter
This User Manual contains information on design, principle of operation, characteristics of the
product and instructions essential for correct and safe use of this product (intended use, maintenance,
servicing, storage and transportation), as well as information regarding the utilization of the product.
1 DESCRIPTION AND OPERATION OF THE PRODUCT
1.1 Product functionality
Dosimeter DKG-07D «Drozd» FVKM.412113.026 (hereinafter - dosimeter) is intended for
measurement of:
-
ambient equivalent dose rate
-
ambient equivalent dose
The dosimeter can be used at nuclear plants and facilities, at radiochemistry plants, in industries
where radiation sources are used, at special and custom control stations, and by ecological services and
sanitary inspections. The dosimeter can also be used by persons from population for individual
monitoring of radiation environment.
1.2 Technical characteristics
1.2.1 Energy range of measured gamma radiation ..................................... 0.05 to 3.0 MeV.
1.2.2 Measurement range:
-
AEDR of gamma radiation ................................................... 1·10
-
AED of gamma radiation .............................................................. 1 to 2·10
1.2.3 Limits of the permissible basic relative measurement error of AEDR and AED of
gamma radiation ................................................................................. (15+2.5/ ) %;
where H is a dimensionless unit, numerically equal to measured values of measured AEDR or AED in
-1
μSv·h
and μSv, accordingly.
1.2.4 Energy dependence relative to
1.2.5 Anisotropy of dosimeter ..................................................................... 30 %:
-
for energies 0.662 and 1.25 MeV and with incidence angle changing from 0° to ±180°
relative to the direction used during dosimeter calibration, in vertical and horizontal planes; except the
angle 90° in horizontal plane, for which the anisotropy of sensitivity is ......not more than minus 45 %;
-
for energies 0,06 MeV with angle changing from 0° to ±45° relative to the direction used
during dosimeter calibration, in vertical and horizontal planes.
1.2.6 Warm-up time of the dosimeter ................................................. not more than 5 s.
1.2.7 Continuous operation with one set of batteries ........................................ 200 hours.
1.2.8 Instability of readings during 8 hours of continuous operation relative to average value of
readings for this period of time .............................................................................. ±3 %.
1.2.9 Interval of measurement of AEDR ..................................................... not limited.
Not e – In the mode of AEDR measurement the precision of readings is continuously improved in the course
measurement. Simultaneously the decreasing value of statistical uncertainty is shown on the LCD that allows to regard the
measurement result as final one when the desired precision is reached.
1.2.10 Power supply voltage ................................................................. 2.0 to 3.2 V.
1.2.11 Power supply of the dosimeter is carried out from two batteries of AA type, 1.5 V each,
with total voltage not more than 3.2 V.
1.2.12 Operating conditions:
-
operating temperature range ..................................................... minus 20 to + 50 C;
-
maximum limit of relative humidity ................................................. 90 % at +25 C;
-
atmospheric pressure .............................................................. 84.0 to 106.7 kPa.
*
Н
10
of gamma radiation (hereinafter - AEDR);
*
Н
10
of gamma radiation (hereinafter - AED).
137
Cs (0.662 MeV) ...................................... 25 %.
FVKM.412113.026RE
3
-1
3
-1
to 3·10
μSv·h
;
5
μSv.