5) MAINTENANCE

WARNING
To avoid electrical shock, disconnect the meter from any circuit, remove the test leads
from the input jacks, and turn OFF the meter before opening the case. Do not operate
with open case.
Trouble Shooting
If the instrument fails to operate, check batteries and test leads etc., and replace as
necessary. Double check operating procedure as described in this user's manual
If the instrument voltage-resistance input terminal is subjected to a high voltage
transient (caused by lightning or switching surge to the system) by accident or
abnormal conditions of operation, the series fusible resistors will be blown (become
high impedance) just as standard fuses to protect the user and the instrument. Most
measuring functions through this terminal will then be open circuit. The series fusible
resistors and the spark gaps should then be replaced by qualified technician. Refer to
the LIMITED WARRANTY section for obtaining warranty or repairing service.
Cleaning and Storage
Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives
or solvents. If the meter is not to be used for periods of longer than 60 days, remove
the batteries and store them separately