Amprobe AMP-25-EUR Процедура калибровки - Страница 2

Просмотреть онлайн или скачать pdf Процедура калибровки для Испытательное оборудование Amprobe AMP-25-EUR. Amprobe AMP-25-EUR 7 страниц.

Amprobe AMP-25-EUR Процедура калибровки
AMP-25, AMP-25-EUR
Calibration Procedure
Periodic calibration is required to conduct correct measurement. This instrument needs
periodic calibration at intervals of one year. Calibrate the instrument according to the following
procedures.
Cautions
 Read instruction manual and understand the use of the instrument and its accessory(ies)
correctly before proceeding the calibration activities.
 Calibrate the instrument at
instrument to sit at this environment condition for > 30 minutes before calibration.
 Erroneous calibrating operation may lead to the incorrect performance and
measurement.
WARNINGS!
 Only trained personnel can do adjustments and repairing of the instrument!
 Be aware of all safety precautions to avoid electric shock when working on opened
instrument!

Recommended equipment:

 Fluke 5520 calibrator
 Transconductance Amplifier
 Current coil with overall diameter < 22 mm

1. Enter Calibration Mode

1.1.
Press and hold the
display symbols, then release buttons and press buttons
following sequence to enter the
LPF
I
inrush
The
"HOLD"
symbol on display will flash when in calibration
1.2.
Press
LPF
I
inrush button, the display shows "A1", press
to select "25"
2. Calibrate the position error (when in calibration mode)
2.1. Press
button to select AC A mode
2.2. Clamp the jaw around the current coil. Ensure the jaw is completely close.
2.3. Apply the reference current AC 100 A / 50 Hz, and position the jaw as below, then
record the reading on the display.
Current coil
Rev. 002
C  2
23
o
o
C
and humidity of
"HOLD"
button while powering up. The display shows full screen
calibration
I
ZERO
HOLD
LPF
Page 2 of 7
Procedure for Item Number:
80% R.H. or less.
immediately
procedure mode.
I
lnrush
mode.
LPF
I
inrush button again
4588254, 4588268
Allow the
according to the