Daiichi Electronics HSQT2-93A Руководство по эксплуатации - Страница 9
Просмотреть онлайн или скачать pdf Руководство по эксплуатации для Преобразователь Daiichi Electronics HSQT2-93A. Daiichi Electronics HSQT2-93A 16 страниц. High speed multi power line transducer
2.5 CE marking (Only auxiliary supply DC24V specification product.)
(1) Conformity technical standard
・EMC conformity technical standard
EMI(Emission) EN61000-6-4
EMS(Immunity) EN61000-6-2
・Safety technical standard
EN61010-1 CATⅢ(Maximum circuit voltage, 300V),Pollution degree 2
(2) Specification according to EMC test item
Electromagnetic
EMI
radiation
disturbance
Electrostatic
discharge
Radio-frequency
amplitude modulated
electromagnetic
field
Fast transients
EMS
Surge
Radio-frequency
common mode
Power-frequency
magnetic field
(3) Installation environment
Over voltage Category Ⅲ
EMC test item
Frequency band
30 ~ 230MHz
230 ~ 1000MHz
Contact discharge
±4kV (Charge voltage)
Air discharge
±8kV (Charge voltage)
Frequency
Field strength
80 ~ 1000MHz
10V/m,80% AM(1kHz)
1.4 ~ 2.0GHz
3V/m,80% AM(1kHz)
2.0 ~ 2.7GHz
1V/m,80% AM(1kHz)
Power port :Line to earth ±2.0kV
Signal port:Line to earth ±1.0kV
Power port :Line to line ±0.5kV
Signal port:Line to earth ±0.5kV
Signal port:Line to earth ±1.0kV
Frequency:0.15 ~ 80MHz
Voltage level:10V
80% AM(1kHz)
Frequency:50/60Hz
Field strength:30A/m
Limits
Quasi-peak 10m distance
40dB(μV/m)
47dB(μV/m)
After test,
Output variation
During test:
Output variation
After test:
Output variation
After test:
Output variation
After test:
Output variation
During test:
Output variation
After test:
Output variation
During test:
Output variation
After test:
Output variation
8
Q-208-081
Conformity
technical standard
EN61000-6-4:2001
EN55011:1998
classA,Group1
EN61000-6-2:2005
±0.5%
EN61000-4-2:1995
EN61000-6-2:2005
±20%
EN61000-4-3:2006
±0.5%
EN61000-6-2:2005
±0.5%
EN61000-4-4:2004
EN61000-6-2:2005
±0.5%
EN61000-4-5:2006
EN61000-6-2:2005
±20%
EN61000-4-6:1996
±0.5%
±20%
EN61000-6-2:2005
EN61000-4-8:1994
±0.5%