4. Batteries and accumulators
At works the PH-T is equipped with a 9V battery. This will last approx. 72 hours when
used for simulation (instrument test). When used for measurement (sensor test) or
high resistance test, power consumption is higher.
NOTE
The battery is accessible from the back of the instrument. Observe the safety and
environment regulations for handling, storage, and disposal of batteries.
The PH-T can be used with rechargeable batteries and provides a charging function.
For specifications of connection and current-supply please refer to the technical data.
The time necessary for recharging of a completely discharged accumulator depends
on the type and the capacity. It can be calculated as follows:
Ni-Cd
Ni-MH
ATTENTION The PH-T does not automatically stop the charging process. Take care
CAUTION
10
The instrument does not indicate low battery.
Charging time (h) = capacity (mAh) x 0.07778
Charging time (h) = capacity (mAh) x 0.08333
not to exceed the calculated charging time.
Never try to recharge regular batteries!