OPERATION
1.
Plug power supply into wall outlet and connect it to the CoolCLAVE™ unit on the backside.
2.
Open the door and put the objects you want to sterilize inside the unit. You can stack or put as many tools as you need, but do not overload
unit so sterilizing can be effective.
NOTE: because the CoolCLAVE uses ozone gas, you are able to safely and effectively sterilize the surfaces of most types of materials such as
plastics, metals, paper products, pipet tip boxes, and even agar plates. On the other hand, the CoolCLAVE is not able to sterilize liquids or
solutions.
3.
Close the CoolCLAVE door and press the "Start" button on the face of the unit.
a.
For each press of the Start button, the CoolCLAVE will run an 8 minute sterilizing cycle.
b.
The green light bar next to the Start button will visually indicate the time left for each 8 minute cycle.
c.
An 8 minute cycle is typically sufficient for a few lightly contaminated items. For heavier loads or further cleaning, press the Start button
again to start another 8 minute cycle. It is recommended that in general you run the CoolCLAVE 2-3 cycles for the best cleaning
effectiveness.
d.
During operation internal LED lights will flash and change color to show that the unit is in working mode.
e.
If during an operating cycle you open the CoolCLAVE door, the unit will stop working. Pressing the Start button again, after closing door,
will restart the 8 minute cycle from the beginning.
MAINTENANCE
(a)
The CoolCLAVE has a small filter cartridge that needs occasional cleaning every 3 months. The cartridge is located on the inside of
the unit, towards the back right side. Remove the cartridge by sliding it out of its holding cage and clean it by using air gun or vacuum
cleaner to remove dust particles. Avoid damaging the cartridge's fabric cover by using moderate air pressure or a soft vacuum brush
head. Alternatively, you may just shake the dust off of the filter in an outdoor area, or use a dry soft brush to remove dust and particles.
(b)
Keep you CoolCLAVE unit clean at all times by wiping the interior and exterior surfaces with a moist towel. To avoid damaging the
unit, do not use any detergents or solvents.
(c)
Unplug the power supply from the wall outlet when the CoolCLAVE is in storage or will not be used for extend periods of time.
(d)
Avoid obstructing the back of the unit by leaving a space of about 2 inches between it and a wall surface.
(e)
Avoid damaging the unit by putting heavy objects on top of it.
The purchase price paid for the CoolCLAVE™ grants end users a non-transferable, non-exclusive license to use the unit in a research
laboratory setting, and for sterilizing purposes only as described in this manual. This use license excludes and without limitation, resale,
repackaging, or modification of the unit in any way and without prior notification of and approval by Genlantis.
Under no circumstances shall the CoolCLAVE unit be used on food or drink products intended for human or animal consumption. At all
times, care and attention should be exercised in handling and using the CoolCLAVE unit by following the instructions in the manual, using
common sense lab practices, and wearing protective lab clothing and eyewear.