Under heating condition, press [star], and the green indicator
lighten, heating timing start when temperature reach the setting
data, machine will alarm start at last 10 seconds.
7. How to make PPP autologous
Step 1: Blood collection
Step2 : Blood separation
Step 3: PPP collection
Step 4: Heat PPP at 80 ℃ for 4 ~8minutes.
(Note:please pre-heat to 80 ℃ before put syringes into)
Step 5:
Put PPP into cooling holes for 5 minutes. , the the PPP
gel make finished.
Step 6: Inject the PPP gel into the needed place
8. Trouble shooting
Please check the problem according to the following table, if still
can not eliminate the trouble, please contact the buyer.
Display
No display on
control board
E-05
Trouble
No voltage or
over-current protection
Temperature sensor
error or trouble in
Measures
Check the power voltage
or change the fuse
Change the temperature
sensor
6